Steady load · Appliances
CPAP machine
ResMed AirSense 11 power from the Americas user guide and ResMed's lab table, from 7 W with no humidifier to a 73.2 W peak.
- Load
- Steady load
| Figure | Value | How it was reached |
|---|---|---|
| Running watts | 7–56.1 W | The low and high figures in the sources. Not averaged into a single "typical" watt. |
| Peak watts | 73.2 W | A short peak printed by the manufacturer. Compared with the station's rated watts, not with a motor surge. |
This page uses ResMed’s AirSense 11 because ResMed prints the watts. A Philips DreamStation or another brand can be different. Do not paste these numbers onto a machine ResMed did not measure.
Two ResMed documents, read together
The Americas user guide specifies a 65 W power supply. AC input is 100–240 V, 50–60 Hz, 2.0 A. Typical power consumption is 56.1 W (111.5 VA). Peak power consumption is 73.2 W (137.6 VA). The pairing check uses 73.2 W against the station’s rated watts. The runtime range uses 7 W at the low end and 56.1 W at the high end.
The second document is ResMed’s power table, article 1018528/11. The operation column is in watt-hours. For an hour of therapy, watt-hours equal average watts. It lists:
- AirSense 11 AutoSet, no humidifier: 7
- With HumidAir 11: 19
- With HumidAir 11 and ClimateLineAir 11: 55
The note on that table says the figures were measured in a lab and that other settings can differ. A plug-in meter at your pressure is the check that beats both documents.
The 19 W and 55 W rows sit inside the 7 W to 56.1 W range the calculator uses. They are cited so a heated-tube night is not invisible between the two ends.
What changes the night
Humidifier heat and a heated tube are the large steps in ResMed’s own table. Altitude, leaks, and pressure are the things the table tells you to measure locally. None of that is a duty cycle. A CPAP that is on for the night is a steady load for those hours. Do not divide by a duty cycle unless you have turned the machine off part of the night, which is a therapy decision and not a battery trick.
Medical limit, said once
A runtime is not a prescription and not permission to travel without the battery or backup your supplier already specified. Jackery’s own FAQ says not to use Jackery stations for medical devices. Other brands may market CPAP use. Marketing is not a clearance. Test the exact machine, on the exact station, for a full night you can afford to lose, before you need it.
Pairings that use this appliance
Questions people ask
How many watts does a CPAP use?
ResMed's lab table lists an AirSense 11 AutoSet at 7 Wh in operation with no humidifier, 19 Wh with a HumidAir 11 humidifier, and 55 Wh with that humidifier and a ClimateLineAir 11 heated tube. The Americas user guide lists typical power consumption of 56.1 W and a peak of 73.2 W for the 65 W power supply. A watt-hour per hour of operation is an average watt.
Is the 65 W power brick the number to use for runtime?
The brick is the supply rating, 100–240 V input and a 24 V output, with typical consumption listed at 56.1 W and peak at 73.2 W. The lab table is lower for the same machine when the humidifier and heated tube are off. Use the setup you actually sleep with. The brick rating alone is a ceiling, not the night's average.